Displaying report 1-1 of 1.
Reports until 11:31, Thursday 27 July 2023
GQuEST General (GQuEST)
Hudson Alexander Loughlin - posted 11:31, Thursday 27 July 2023 (11256)
Profile of Beam Headed to Indium Cavity
We took a profile of the beam after the faraday headed to the cavity. To compute the beam parameters, I fit parameters to a TEM00 mode instead of using the beamspotsize python package to calculate parameters since that package allows M^2 to float, and the beam waist is inside the Faraday and unaccessible. Taking the folding mirror after the faraday to be z = 0, the waist is located at z = -102.5mm +- 26.1mm and the waist spot size is w0 = 55.0um +- 5.1um. The attached figure shows the curve fit results.
Images attached to this report
Displaying report 1-1 of 1.