Displaying report 1-1 of 1.
Reports until 16:57, Friday 09 February 2024
GQuEST Output Filter Cavities (GQuEST)
Daniel Grass - posted 16:57, Friday 09 February 2024 - last comment - 17:00, Friday 09 February 2024(11345)
Noliac NAC2125-H08 (Ring Piezo) Testing

Alex and Daniel

We checked a Noliac NAC2125-H08 for shorts and found none. We measured the capacitance to be 2.4641 uF. The expected capacitance is 2.4 uF. We also measured a resistance of 3.5082 kOhm. When we pressed on the piezo, we noticed a voltage. We flagged one of the wires for polairty. See attached photos. We also noticed that the nylon tipped set screws that hold the piezo in place provide a voltage.

Images attached to this report
Comments related to this report
Daniel Grass - 14:34, Thursday 26 October 2023 (11354)

I looked for the dots on the Piezo we tested, and they are on the wire we DID NOT FLAG. I think we should flip which wire is flagged and then use that convention going forward.

Daniel Grass - 17:00, Friday 09 February 2024 (11355)

I flagged the lead of the piezo with the black dot, which is explicitly stated by Noliac as the positive electrode. When I press on the piezo, there is a positive voltage when measuring off this positive lead.

Images attached to this comment
Displaying report 1-1 of 1.