GQuEST Output Filter Cavities
Torrey Cullen - posted 12:23, Friday 05 January 2024 (11405)
Modematching Success

I added a lens between the two EOMs to ensure no clipping on the second EOM. I then reprofiled and found a MM solution to put the waist of 600 um at M3 (3.003 m from colimator output) in the cavity. This has been implemented and measured to have an average beam waist and location of waist (average of x and y, beam is slightly astigmatic) of 599.5 um @ 2.97 m. This should more than suffice to see some amount of resonance in the cavity. The nextcloud server is currently down but "Profiling after l1 12072023.ipynb" has this data on the lab computer. Will also update the layout mockup when the server is back up.