[Torrey, Sander, Erin, Briana]
General notes from initial tests of silicon wafer mirrors.
-They came out of the box with visible amounts of dust on them.
-We need a mirror mount that allows for small thickness mirrors. Using the colored mirror mounts in the mean time.
-Turn power down at power distribution to 10 mW coming out of the MISC path. Use this for measuring 1550.
1550 on the wafer mirror
-10.2 mW input / 5.6 mW tranmission - normal incidence
-10.2 mW input / 5.8 mW reflection - 45 AOI
775 on the wafer mirror
-3.5 mW input / transmission too small to measure - normal incidence
-3.5 mW input / 1.0 mW reflection - 45 AOI