Jeffrey Wack - posted 18:11, Monday 01 July 2024 (11696)
ULN15TK laser DC voltage control calibration

Here are some calculations regarding the DC control of the frequency of the Thorlabs ULN15TK laser in B102. The measured value of the Hz/V laser tuning coefficient disagrees with the value calculated from the datasheet by roughly a factor of 2.

 

First, the laser frequency was swept with a 2Vpk-pk triangle wave at 10Hz. Two resonance peaks were identified to be 30ms apart. We want to convert this time difference into a voltage difference using the properties of the tringle wave.

30ms*2V*2*10Hz = 30ms*40V/s = 1.2V

The free spectral range of the cavity, measured in Hz is the inverse of the travel time in the cavity, or c/L

FSR = 3e8/2.4 = 125MHz

Together these values give 104MHz/V

 

We should also be able to calculate this value (Hz/V) using only the datasheet of the laser.

Under the specs tab, it is reported that the DC voltage to current conversion rate is 2 mA/V, and the current tuning coefficient is 0.25pm/mA

This gives us 0.5pm/V tuning. We now convert this to frequency.

0.5pm/V*(3e8m/s)/(1550nm)^2 = 62MHz/V

 