Environmental Dust/Particulates
Daniel Grass - posted 12:51, Monday 08 July 2024 - last comment - 16:11, Monday 08 July 2024(11715)
Particle Counter Measurements

I took some particle counter measurements in B102 and some LIGO optics labs in Lauritsen. x/x/x is the 0.3 um, 0.5 um, and 1.0 um cumulative count over 1 minute.

B102:

77973/4322/471 before fans were turned on. The fans were left off over the weekend.

9/0/0 after the fans were on for 90 seconds before starting the count

 

LIGO clean room to measure the scatter off of our optics, including the super-polished optics:

1/0/0 near curtains

0/0/0 2 feet inside curtains

 

Room with Fizeau Interferometer

84868/3495/397

GariLynn Billingsley says they clean the LIGO test masses a lot after they are in this room, but this is definitely not a clean room

Comments related to this report
Ian MacMillan - 14:57, Monday 08 July 2024 (11716)

Were all of the fans off? We left the fans on our table (the larger one) on when we left. Were They both off?

Daniel Grass - 16:11, Monday 08 July 2024 (11717)

Those fans were on