Displaying report 1-1 of 1.
Reports until 18:02, Thursday 05 September 2024
Electronics General
Jeffrey Wack - posted 18:02, Thursday 05 September 2024 (11859)
Some laser modulation noise tests
coupling impedance present in pdh error?
AC 50Ohm -83dBm
DC 50Ohm -68dBm
AC 1MOhm -86dBm
DC 1MOhm -86dBm
Unplugged not present

I connected the laser DC modulation port to the Moku input of a spectrum analyzer. The spectrum analyzer always sees the 875Hz if it is connected to the laser, but the amount that the 875Hz shows up in the pdh error spectrum changes a bit on changing the spectrum analyzer input properties. Could the problem have something to do with impedance matching? the input impedance to the laser modulation port is 1kOhm.

We should plug the laser modulation port into a non-Moku spectrum analyzer and see if the problem persists.