Reports until 18:15, Thursday 19 September 2024
Environmental Dust/Particulates
Daniel Grass - posted 18:15, Thursday 19 September 2024 - last comment - 14:48, Monday 13 January 2025(11873)
Particle Count Measurements in B102, B111, B111A, and B111B

I took some particle counter measurments in B102, B111, B111A, and B111B. In all the clean rooms, the particle count was 0. These clean rooms were the B102 cleanroom with the optical cavities, the B111B mobile cleanrooms, the Southern most B111A long table, and the Northern most B111B short table.

I also took some 10 second particle counts outside the clean rooms. Data is 0.3 um count/0.5 um count/1.0 um count)

B102: 1268/72/28 (there were lots of people and activity during this measurment)

B111 (entryway): 495/46/28

B111B: 177/51/39

B111A: 95/17/13

These seem like very good measurments.

The particle counter is on the optics tables in B111B.

Comments related to this report
Daniel Grass - 19:58, Sunday 12 January 2025 (12070)

It should probably read "Northern most B111A short table" considering there is no short table in B111B.

Daniel Grass - 14:48, Monday 13 January 2025 (12072)

Data extrapolated to a full minute for easy comparisons:

B102: 7608/432/168 (there were lots of people and activity during this measurment)

B111 (entryway): 2970/276/168

B111B: 1062/306/234

B111A: 564/102/78