JC and I picked the lock of the locked lista cabinet in the cryo lab and got the IR viewer. I used it to observe the power distribution center. I did not observe any significant amounts of scattered 1550nm light. There was a fair amount visible on the SHG sled where irises are being used to block the 0th order beam but this is 775. The amount of power there is known to be low though, I don't think this is an issue.
In short, I'm not sure we need one of these viewers. I could not observe any scatter where I expected there to be some. It could easily see 775 nm, as expected as they are several orders of magnitude more sensative at this wavelength. One caveat potentially is this looks like a much older model than what is currently being sold, maybe the technology has improved since this older version.