Vacuum Systems General (GQuEST)
Daniel Grass - posted 17:03, Wednesday 18 December 2024 (12060)
10 in CF Flange to Rotatable 8 in CF Flange 12 in long Custom Nipple Installed on Laser Filter Cavity Output Vacuum Cube

[Jeff, Daniel]

We attached the 10 in CF Flange to Rotatable 8 in CF Flange 12 in long Custom Nipple onto the Laser Filter Cavity (LFC) Output Vacuum Cube. When dropping in the copper gasket, we used 12-14 bottom bolts instead of the usual 4 because the nipple applied a large torque pulling the top away from the cube. Otherwise, we installed it as usual and I tightened the bolts to 34 Nm.