Environmental Dust/Particulates
Daniel Grass - posted 14:00, Tuesday 14 January 2025 (12074)
Particle Count Measurements on January 14, 2025 in B111, B111A, B111B

I took some particle counts today with the full 1 minute capture time. They are much better than yesterday.

B111A North large table North small table South large table South small table Outside clean rooms in middle of room south of curtain Outside clean rooms west of north small table before air purifier was turned on  Outside clean rooms west of north small table after air purifier was turned on for ~7 minutes
0.3 um 9 1 411 1 56 145 81
0.5 um 2 0 50 0 19 52 26
1.0 um 0 0 13 0 12

34

17

See attached photo for the location of the air purifier (with a newly installed filter and run at fan setting 3 with UV on) and the particle counter for the last two measurements.

 

B111B Control area Outside clean room by northeast shelves Inside clean room
0.3 um 3526 77 1
0.5 um 1598 48 0
1.0 um 579 26 0

The 0.3 um particle happened within the first second of the minute capture time, so might be been caused by me setting it up (I didn't have a beard net or gloves on).

 

Other B111 antechamber
0.3 um 1516
0.5 um 489
1.0 um 235

 

For each respective location outside the clean rooms, these are some of the lowest particle counts we have taken. Particle counts are a bit elevated inside the clean rooms.