Torrey Cullen - posted 15:16, Tuesday 14 January 2025 (12075)
First Cavity Lock post 2025 Eaton Canyon Fire

We had been discussing if the extremely poor air quality due to the fire had effected our superoptics in the OFCs. To check this I performed a couple quick ringdown measurements on OFC1. For both wavelengths, the finesse of the cavity is close to measurements before the fires. It seems uneffected. For reference F = pi * c / L * (measured tau in screenshot).