Reports until 16:50, Tuesday 04 March 2025
Environmental General
Jeffrey Wack - posted 16:50, Tuesday 04 March 2025 (12160)
Lab boot up post repairs

[Torrey, Jeff]

After the conclusion of the drywall repairs, we turned back on the lab electronics. We

Before turning on the lasers we:

We then tested the power output of both lasers on each end of the 50:50 beamsplitter in the drawer.

  input to amplifier pickoff path
Thorlabs 22.8mW 23.6mW
Teraxion 18.5mW 19.6mW

These power levels are satisfactory, each laser can meet the laser amplifier's power threshold.

We then turned on the laser amplifier and measured 460mW after the faraday isolator on the power distribution sled.