Displaying report 1-1 of 1.
Reports until 21:12, Wednesday 09 April 2025
Vacuum Systems General (GQuEST)
Daniel Grass - posted 21:12, Wednesday 09 April 2025 (12235)
Vacuum Tubing Connected from Laser Filter Cavity Turbo Pump to Scroll

I connected the ~2 ft long KF 16 to KF 25 tubing from the Agilent TwisTorr 74 FS Turbo Pump to a 4 ft long KF 25 tubing which was connected to the Agilent IDP 7 scroll pump. I used the included filter inside the o-ring going to the scroll pump.

Now that the vacuum system is fully connected, I opened up the gate valves to the turbo pump and the ion pump. I also connected the power cord to the scroll pump but didn't plug it into a wall. I flipped a red switch so that the scroll pump is expecting 110-125 V instead of 230 V.

Images attached to this report
Displaying report 1-1 of 1.