[Ian, Jeff, Daniel]
We connected the IDP 7 Scroll Pump for the Laser Filter Cavity (LFC) to the same power system as the SNSPD Dewer pumps. We connected the Turbo and Turbo Controller to a power strip on the top shelf in the mobile clean rooms. The pressure gauge, after warming up for a few minutes, read "pressure too high". This is definitely a measurement error, but I think it's fine. I plugged in the USB-B extension cords from amazon and they successfully power the fan that cools the turbo pump.
We turned on the scroll pump at 4:24 PM, and after ~30 seconds it made a secondary "turning on" mechanical sound and the pressure gauge started to read actual values instead of "pressure too high". After 20 minutes, the pressure was 0.1 Torr. We plugged in the interlock bypass D-Sub and turned on the turbo. After 2 minutes (and 1 vertical bar on the turbo controller), the pressure was 3*10^-4, after 12 minutes, the pressure was 8.7*10^-6 Torr. The 3 vertical bars then revealed the turbo speed, 1167 Hz. The final power was 12 W after peaking around 90 W while ramping up. I left after 15 total minutes of the turbo being on with a pressure of 6.9*10^-6 Torr.
Pressure at 6:53 PM last night: 1.2E-6 Torr
Pressure at 11:30 AM today: 2.8E-7 Torr