[Torrey,Jeff,Ian]
We decided to check on air quality in bridge/lab. The hallways and offices are suprisingly decent. As you enter the lab it is more and more smokey. Inside the cleanroom in B111B is by far the worst smelling. May need to replace the filters in the cleanroom fans. For now though, cleanroom1.jpg and cleanroom2.jpg show no particles above .3 um in the cleanroom. Outside the cleanroom the particle counts seem higher than normal, would have to find old log posts with these numbers though.
Additionally we did a quick sweep in B111A. It smells like an animal has died in the vents somewhere. I will look into putting in a maintenence request tomorrow morning. Particle count where the smell is the strongest: deadanimal.jpg. Additionally the B111A cleanroom fans do not seem to be on. insideB111Acleanroom.jpg (this is on the table with the RQB stuff).
From September when we were moving in: https://mccullerlab.com/logs/lab/index.php?callRep=11873
These results are ~5x worse in B111B outside the clean room and ~30x worse in B111A outside the clean room. There could be some variance in room location, but surely not 30x unless the particle counter was in a really weird location or there was a lot of actively right before and during the count.